SPXCLCAL Calibration / flow cap
Calibration / flow cap. Accessories are outside the electrical-safety and gas-performance certificate.

Ordering data
| P/N | SPXCLCAL |
|---|---|
| Type | Accessory |
| Name | Calibration / flow cap |
| Series | Sensepoint XCL |
| Note | Accessories are outside the electrical-safety and gas-performance certificate. |
Photos and figures


Related part numbers
Install / replace
Fit the calibration/flow cap over the sensor port and apply test gas (~300 ml/min; CO2 ~500 ml/min). It is not a detector.
FAQ
- Is SPXCLCAL a controller?
- No. XCL is a fixed detector. For a controller see Touchpoint Plus (product-4012).
- How is it different from Sensepoint XRL?
- XRL is another family (product-9015): aluminium IP66 with ATEX/IECEx/cULus. XCL is polycarbonate Safe Area IP65. Bluetooth SIG lists both, but housings and certificates are not interchangeable.
- Does it have ATEX?
- No. XCL is certified as Safe Area electrical safety and gas performance — not ATEX/IECEx and not China fire CCC. Certificates are blurred thumbs only, no download.